9.1 AC VOLTAGE
range
2 V
20 V
200 V
600 V
Impedance: 10 MΩ, max. input voltage: 600 V DC or AC rms
Frequency range: TRMS 40 Hz-1000 Hz
9.2 DC VOLTAGE
range
200 mV
2 V
20 V
200 V
600 V
Impedance: 10 MΩ, max. input voltage: 600 V DC or AC rms
9.3 AC CURRENT
range
2 A
20 A
200 A
600 A
Frequency range: TRMS 40 Hz-1000 Hz
V. 04 – 25/02/2020
DCM120
Do not measure circuits that may contain voltages
resolution
0.001 V
0.01 V
0.1 V
1 V
Do not measure circuits that may contain voltages
resolution
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
Do not measure circuits that may contain voltages
resolution
0.001 A
0.01 A
0.1 A
1 A
> 600 V AC or DC
> 600 V AC or DC
> 600 V AC or DC
9
accuracy
± (1.0 % + 5)
± (1.2 % + 5)
accuracy
± (0.5 % + 3)
± (0.8 % + 5)
± (1.0 % + 5)
accuracy
± (2.5 % + 10)