Fault – Cause – Remedy
9
Fault – Cause – Remedy
Errors indicated by the analyzer
Faults or errors, their possible causes and how to remedy them are listed below in alphabetical
order. Repeat the measurement after remedying the fault or error or call DrägerService, see fax
form on page 65.
Message
Signal error
"Display frozen"
Barcode error
Chip too fast
Chip too slow
Chip unevenly
Remove chip 0
Battery empty
Chip error
Remove 0
Dräger CMS Analyzer
Cause
Moisture has condensed
inside the analyzer.
Chip is dirty.
Fault in electronics/optical
system.
Internal fault, ESD,
electrostatic discharge
Barcode on chip cannot be
read.
Batteries are completely
exhausted, measurement is
no longer possible.
Chip type cannot be
processed.
Remedy
Switch off analyzer: move sliding
switch to position 0, remove chip.
Wait approx. 5 minutes for
temperature balance to be
established, then repeat
measurement.
Switch off analyzer: move sliding
switch to position 0, remove chip.
Wipe chip with a dry cloth or insert
a new chip. Repeat the
measurement.
Switch off analyzer:
move sliding switch to position 0,
wait at least 1 minute then switch
on again. Call DrägerService if the
fault cannot be remedied in this
way.
Switch off analyzer:
move sliding switch to position 0,
wait at least 1 minute then switch
on again.
Call DrägerService if the fault
cannot be remedied in this way.
Eject chip: move sliding switch to
position 0.
Check condition of chip and clean if
necessary: wipe chip with a dry
cloth.
Move sliding switch to position 1,
insert chip evenly.
Change batteries, page 40.
Eject chip: move sliding switch to
position 0 and repeat
measurement.
59