Operation
The
chip
parameters
alternately once it has been inserted.
(Example).
Gas to be measured:
NH3
Two texts are displayed alternately in the
bottom line:
Measuring range:
2 – 150 ppm
AND
the prompt to move the sliding switch to
position 2.
Example:
start
2 =
7.4

Chip integrity test

1. Move sliding switch to position 2.
The measurement channel is checked,
prepared for the measurement and
cancelled by pressing the cancellation
pin into the chip.
2. Wait until chip o.k. is displayed. The user
is simultaneously prompted to move the
sliding switch to position 3 (measurement
position).
Example:
chip o.k. = Chip is functional
start measure 3 ↓ = Move the
sliding switch to position 3 when you
are ready to start the measurement.
NOTICE
i
i
Do not delay for too long before moving from position 2 to position 3!
Dräger CMS Analyzer
are
displayed
Move sliding switch to
position 2 to start the
chip integrity test.
NH3
2 – 150 ppm
NH3
start
2
chip test
please wait
chip o.k.
start measure 3
03222837.eps
03322837_4en.eps
03422837_4en.eps
03522837_4en.eps